Tech Specs

Pro AED Tech Specs

DDU-2400 Defibrillator

TYPE
Semi-Automatic external defibrillator 
with Manual Override

MODELS
DDU-2400 series

WAVEFORM
Biphasic Truncated Exponential
(Impedance compensated)

ENERGY
AED Mode -
  Adult: 150 Joules
  Child / Infant: 50 Joules
  (Nominal into 50 Ohm load)
Manual Mode -
  User selectable from
  25-200 Joules
  (Nominal into 50 Ohm load)

CHARGE TIME (new, at 25º C)
Less than 4 seconds
(shock decision to shock)

DISPLAY
High-resolution color LCD

VOICE PROMPTS
Full motion video
On-screen text prompts
Extensive voice prompts guide user through operation of the unit

CPR COACHING
Video and voice coaching
On-demand video help

CONTROLS - BUTTONS
1. Lighted On/Off button
2. Lighted Shock button
3. Menu Selected buttons

RESCUE PROTOCOL
AHA/ERC 2010
Supports protocol updates by the user
(password protected)

WARRANTY
8 year Warranty


Self Tests

AUTOMATIC
Automatic daily, weekly and
monthly circuitry tests

BATTERY INSERTION
System integrity test on battery
insertion

PAD PRESENCE
Pads pre-connected tested daily

USER-INITIATED
Unit and battery pack system
test initiated by the user

STATUS INDICATION
Visual and audible indication of unit status

STATUS SCREEN
Unit self-test results
Pads and battery information
(status and expiration)


download tech specs pro aed defibrillator

Patient Analysis System

PATIENT ANALYSIS
Automatically evaluates patient impedance for proper pad contact. Monitors signal quality and analyzes patient ECG for shockable/non-shockable rhythms

SENSITIVITY/SPECIFICITY
Meets AAMI-DF-80 specifications and AHA recommendations


Defibrillation / Monitoring Pads

MODEL
Adult - DDP-2001
Child / Infant - DDP-2002

TYPE
Pre-connected, single-use, non-polarized, disposable, self-adhesive electrodes with cable and connector

SURFACE AREA
77 cm2 (nominal, each pad)
50 cm2 (nominal, each pad)


Environmental

TEMPERATURE
Operating: 0 to 50°C
Standby: 0 to 50°C

RELATIVE HUMIDITY
Operating / Standby: 5% - 95%
(non-condensing)

ALTITUDE
-500 to 15,000 ft (-150 to 4500m) per MIL-STD-810F 500.4 Procedure ll

VIBRATION
Ground (MIL-STD-810F 514.5 Category 20)

Helicopter (RTCA/DO-160D, Section 8.8.2, Cat R. Zone 2, Curve G)

Jet Aircraft (RTCA/DO-160D Section 8. Cat H, Zone 2, Curves B & R)

SHOCK / DROP ABUSE TOLERANCE
MIL-STD-810F 516.5 Procedure IV 48 inches (1.2 meters), any edge, corner, or surface, in standby mode

CRUSH TEST
1,000 pounds (450 kg)

SEALING / WATER RESISTANCE
IEC60529 class IP55; Protected against dust and waterjets (battery pack installed)

ESD
EN61000-4-2: (Open air up to 8kV or direct contact up to 6kV)

EMC (Emission)
CISPR 11 Group 1 Level B
and FCC Part 15

EMC (Immunity)
IEC 61000-4-3 and IEC 61000-4-8

Battery Pack Options

MODEL DBP-2003

POWER
12V, 2800 mAh

CAPACITY (new, at 25º C)
125 shocks or 8 hours continuous operation

STANDBY-LIFE (typical)
4 years

TYPE
Lithium/Manganese Dioxide Disposable, recyclable, non-rechargeable

LOW BATTERY INDICATORS
Visible
Audible


Event Documentation

INTERNAL EVENT RECORD
Critical ECG segments and rescue event parameters are recorded (greater than 60 minutes) and can be downloaded to a removable data card

PC-BASED EVENT REVIEW
ECG with event tag display, and audio playback when available

REMOVABLE STORAGE
(optional) Up to 30 hours of ECG and event data storage (no audio option) or up to 3 hours of audio (audio option). ECG and event storage on a removable data card. Actual length of storage is dependent on card capacity

USB PORT
Event download and maintenance operations


Event Documentation

SIZE
18.5 x 24 x 5.8 cm

WEIGHT
Less than 1.4 kg
(with battery)


Operating Modes

AED MODE
With Video Display (animated instructions and CPR coaching) or with ECG Display

MANUAL MODE
Displays high-resolution ECG data and event information and defibrillation energy selection from 25-200 Joules with charge, shock and disarm control

ECG MONITOR MODE
Displays high-resolution ECG data and event information (with 3-lead electrodes)